Switching Mobile Hearing Testing Vendors Without Losing Your Baselines

November 19, 2026 · 6 min read

By Jeff Schroeder, Foster Special Instruments

Does switching mobile hearing testing vendors reset employee baselines?

No. A baseline audiogram belongs to the employee's hearing conservation history, not to the company that ran the test. Under 29 CFR 1910.95(g)(9), an annual audiogram may replace the baseline only when the audiologist, otolaryngologist, or physician reviewing it judges that a standard threshold shift is persistent, or that the annual thresholds show significant improvement over the baseline. Nothing in that paragraph lets a new vendor start fresh. OSHA has applied the same logic to a harder case. In a March 2011 letter of interpretation, it said a successor employer that buys a plant generally may not establish new baselines and should continue to rely on the former employer's. If a new owner cannot reset baselines, a same employer changing vendors cannot either. The new vendor's first test is an annual audiogram, compared against the original baseline.

Who owns the audiogram records, the employer or the vendor?

The employer. Paragraph (m)(2)(i) of the noise standard says the employer shall retain all employee audiometric test records, and (m)(3)(ii) says those records are kept for the duration of the affected employee's employment. A mobile testing vendor may store the files in its software, but that is a service arrangement, not a transfer of the legal duty. The employer also answers requests for those records. Paragraph (m)(4) applies the access rules in 29 CFR 1910.1020, which require access in a reasonable time, place, and manner, and if access cannot be provided within 15 working days, the employer must explain the delay and give the earliest available date. An employer that lets its only copy of the audiogram history leave with a former vendor cannot meet that obligation. The practical rule is simple: before the contract ends, the employer should hold a complete copy of every record, in a format someone other than the old vendor can read.

What records should you get from the old vendor before switching?

Ask for the full history, not just last year's results. At minimum, the record set should cover what 1910.95(m)(2) requires for each audiogram, plus the program documents that explain it.

RecordWhy it matters after the switchRule
Baseline audiogram for every enrolled employeeEvery future STS comparison runs against it1910.95(g)(5), (g)(10)
All annual audiograms, including revised baselinesShows which baseline is current and why it changed1910.95(g)(6), (g)(9)
Name, job classification, test date, examiner nameRequired contents of each audiometric record1910.95(m)(2)(ii)(A)-(C)
Date of last acoustic or exhaustive calibration of each audiometer usedProves the old tests were valid1910.95(m)(2)(ii)(D)
Employee's most recent noise exposure assessmentRequired with each audiometric record1910.95(m)(2)(ii)(E)
Background sound pressure level measurements for test rooms or the mobile unitProves the test environment met Appendix D1910.95(m)(2)(ii)(F)
STS determinations, reviewer notes, and employee notification lettersExplains open follow-ups and OSHA log entries1910.95(g)(7), (g)(8)

Get the audiograms as threshold values by frequency and ear, not only as PDF printouts. Thresholds at 500 through 6000 Hz are what the new vendor's software needs to run the comparison, and retyping them from images is where transcription errors creep in.

What if the old records are missing or incomplete?

Start by separating missing from invalid. The 2011 OSHA letter says time away from noise does not make a baseline invalid. A baseline is invalid only if it did not meet the testing requirements in 1910.95(h): all required frequencies (500, 1000, 2000, 3000, 4000, and 6000 Hz) in each ear, an audiometer meeting the referenced ANSI S3.6 specification, a test room meeting Appendix D, and calibration per Appendix E. So a baseline with a missing calibration date or a missing frequency is a different problem from one that was simply never handed over. For records the old vendor has not produced, send a written request, keep the reply, and escalate through the contract. If a baseline truly cannot be recovered, the standard does not spell out a procedure. Document the search, and let the reviewing audiologist or physician decide how the next audiogram is treated, rather than having the new vendor quietly label it a baseline.

How should you time the switch around annual test dates?

Annual audiograms are due at least annually after the baseline under 1910.95(g)(6), and a vendor transition is a common way for a few employees to slip past their anniversary. Before the old contract ends, build a list of every enrolled employee with their last test date. The new vendor's first visit needs to land before the oldest of those dates. New hires deserve separate attention. The baseline is due within 6 months of first exposure at or above the action level, or within 1 year when a mobile test van is used, as long as the employee wears hearing protectors from the sixth month until the baseline is done. A hire who started during the transition can fall between two vendors' rosters, so pull that list from HR, not from either vendor. For more on the windows, see our post on baseline audiogram rules.

Does an STS found by the new vendor go on the OSHA 300 Log?

It can. The vendor changed, but the employer did not, so a recordable shift goes on the employer's own log. Under 29 CFR 1904.10(a), a work-related STS is recordable when the employee's total hearing level is also 25 dB or more above audiometric zero, averaged at 2000, 3000, and 4000 Hz, in the same ear. That determination only works if the new vendor compares against the true baseline. A vendor that treats its first test as a baseline will miss any shift that happened under the old vendor, and one that compares against last year's annual instead of the baseline will miss slow loss. Ask the new vendor, in writing, which audiogram its software uses as the reference for each employee. The details are in our posts on calculating a standard threshold shift and STS recordkeeping on the 300 Log.

How Foster handles a vendor switch

When an employer moves its hearing testing to us, the first request we make is for the prior history: every baseline, the most recent annual, and any revised baselines with the reviewer's notes. We ask for it weeks before the first test day, because records from another vendor rarely arrive complete on the first try. Each employee's baseline thresholds go into our system before the unit rolls, so the annual comparison on test day runs against the original baseline, not against our own first audiogram. Employees with no baseline on file get flagged to the employer before testing, not after. We also build the first on-site testing date from the employer's oldest anniversary, so nobody's annual lapses during the handoff. If you are comparing providers, our mobile hearing testing vendor checklist covers what to ask before you sign.

FAQ

Can the new vendor just test everyone and call it a new baseline?

No. Baselines change only through 1910.95(g)(9), on the reviewer's judgment of a persistent shift or significant improvement. A vendor switch is not one of those conditions.

How long must we keep the old vendor's audiograms?

For the duration of each affected employee's employment, under 1910.95(m)(3)(ii). Noise exposure measurement records have a shorter minimum of two years under (m)(3)(i).

Can the old vendor charge for or withhold our records?

The standard puts the retention duty on the employer and does not address vendor contracts. Check the contract's records and termination terms, and get a full export before the last test day.

Do we need the old audiometer calibration records too?

Yes. Each audiometric record must include the date of the audiometer's last acoustic or exhaustive calibration. Without it, the old tests are harder to defend. Our calibration page covers what those checks involve. To plan a switch with the history transfer built in, request a quote.