Replacing Audiometer Headphones: Does It Need Recalibration?

November 26, 2026 · 6 min read

By Jeff Schroeder, Foster Special Instruments

A headset cord fails mid-week. A cushion cracks. Someone drops an earphone on a tile floor, or a tech borrows the headset from the spare audiometer to keep testing. Each of these looks like a small hardware fix, and each raises the same question: does the audiometer need calibration again before the next employee sits down? The answer depends on which part changed. This post sorts the common cases, explains why the calibration record is tied to specific earphones, and covers what to do before testing resumes.

Why is an audiometer's calibration tied to its earphones?

An audiometer does not make sound by itself. The electronics send a signal, and the earphone turns it into sound pressure at the ear. Calibration measures that final sound pressure, so the result belongs to a pair: this chassis with these earphones. The acoustic check in Appendix E of 29 CFR 1910.95 is built on that idea. The earphone is placed on an NBS 9A coupler, and a sound level meter with an octave-band filter set measures sound pressure at each test frequency from 500 to 6000 Hz, for each earphone. The readings are compared with Table E-1 for TDH-39 earphones or Table E-2 for TDH-49 earphones, as appropriate for the type of earphone. Reference levels are therefore specific to the earphone model, and the correction stored in the unit is specific to the transducers that sat on the coupler. Change the earphone and the stored correction no longer describes what reaches the ear.

Which headphone changes require recalibration?

What changedRecalibration needed?What to do before testing
Same-model ear cushion replacedUsually noRun the daily functional check on a person with known, stable thresholds and listen for distortion
Headband or cord replaced, earphone element untouchedUsually no, but a new cord can fail or crackleDaily functional check, then a closer listening check at each frequency
Earphone dropped or struckTreat as yesAcoustic check on the coupler before the unit returns to service
One earphone replaced with the same modelYesAcoustic calibration with the new earphone on the coupler
Different earphone model (for example TDH-39 to TDH-49, or to inserts)YesAcoustic calibration against the reference levels for the new type
Headset moved from another audiometerYesAcoustic calibration of the pair as used, and update the records for both units

The cushion and cord rows carry a hedge on purpose. A cracked or mismatched cushion can still change what a person hears, which is why the functional check comes first.

Is replacing an ear cushion a calibration event?

Usually not. Cushions are a wear item, and cracked foam or vinyl is a common reason a headset needs attention. Swapping a worn cushion for the same model from the manufacturer or a trusted supplier is routine maintenance. What matters is what you do next. Under 1910.95(h)(5)(i), the audiometer's function is checked before each day's use by testing a person with known, stable thresholds and listening for distorted or unwanted sounds, and a deviation of 10 dB or greater requires an acoustic calibration. A cushion swap is a good reason to run that check immediately, in both ears, before the next employee is tested. If the person's thresholds hold, the swap did no harm. If they move, stop and have the unit checked. A different cushion style, such as a thicker noise-excluding type fitted to a standard earphone, is a different case. Ask the headset or audiometer maker before using it on a calibrated unit.

What happens when a headset cord fails or an earphone is dropped?

A failed cord is often just a failed cord. Intermittent static, a dead channel, or a signal that cuts out when the cable is moved usually points to the wiring or the connector, and replacing it with a correct part may not alter the earphone itself. Even so, confirm the repair with the functional check and a careful listen at every frequency before relying on it. A dropped earphone is different. The transducer inside can shift, crack, or lose sensitivity without any visible damage, and the shift is not predictable. A unit with a dropped earphone should go back on the coupler before it tests anyone, because the daily functional check is designed to catch a large shift, not to measure a small one. If the person's thresholds move 10 dB or more, an acoustic calibration is required. If the shift is smaller, you may be testing employees with an error you cannot see. Treat a drop as a reason to measure, not to guess.

Why does swapping headsets between audiometers break the calibration record?

It is tempting to keep testing by borrowing the headset from the spare unit. The problem is that each audiometer was calibrated with its own pair, and the certificate describes that pairing. Move a headset and two things happen. The borrowed headset is now attached to an audiometer whose stored correction was set for different transducers, so the levels at the ear are unverified. The unit that lost its headset has a certificate describing earphones it no longer uses. Every audiogram taken after the swap lists a calibration date, as 1910.95 requires for the audiometric test record, that points to a calibration which did not cover the earphones actually in use. Two earphones of the same model can differ in output, and OSHA's annual check measures each earphone separately for that reason. If a swap is unavoidable, treat both audiometers as needing acoustic calibration and keep a note of which headset was on which unit and when.

Does it matter that a newer standard is used for the calibration?

Not for the headset question. In a 2010 interpretation letter, OSHA stated that it considers ANSI S3.6-2004 equally protective as S3.6-1969, and that using the updated consensus standard is at most a de minimis violation. Under either version the reference values apply to the earphone type in use, so name the earphone on the calibration record. If you are unsure a replacement is the same model, check the marking on the housing or ask the calibration provider before ordering.

How Foster handles this

Foster calibrates audiometers in our shop and on site across Ohio, Kentucky, and Indiana, and we calibrate each unit with the headset it ships with. The certificate records the transducer serial numbers so the pairing is written down. We ask customers to send the headset along with the audiometer, since a calibration without it only tells you about the electronics. We can also supply a replacement headset and calibrate it to your unit before it goes out. If a cord fails between visits, call before you swap parts, because a short description is often enough to tell whether a coupler check is needed. See our calibration service, on-site testing for customers who cannot lose a testing day, and hearing testing. To get a headset or a unit checked, request a quote.

For more on what an out-of-spec unit looks like, read signs an audiometer is drifting out of calibration. If the unit has to travel, how to ship instruments for calibration without damage covers packing the headset so it arrives intact.

Frequently asked questions

Can I use any TDH-39 or TDH-49 headset on any audiometer?

Physically it may plug in, but the calibration will not match. The unit was calibrated with specific earphones, so a different headset needs its own acoustic calibration on that audiometer before it is used for testing.

Do I need to recalibrate after changing only the foam or vinyl cushion?

Usually not for a same-model cushion. Run the daily functional check right away and listen for anything unusual. If thresholds shift by 10 dB or more, an acoustic calibration is required.

What if the headset was dropped but it still sounds fine?

Sounding fine is not a measurement. Have the earphone checked on a coupler before testing employees, since damage inside the transducer is not always audible or visible.

Do insert earphones use the same reference levels as standard headphones?

No. Insert earphones have different reference levels than supra-aural ones, so changing the transducer type means calibrating the audiometer against the right reference for the new type.