Overdue Audiometer Calibration: What Happens to the Tests?

November 18, 2026 · 6 min read

By Jeff Schroeder, Foster Special Instruments

Calibration dates slip. A tech leaves or the reminder lands in a shared inbox nobody reads. Then annual hearing tests start, and a few weeks in someone spots that the last calibration certificate is fourteen months old. The questions come fast: are the audiograms we just ran valid, do employees need to be retested, and what do we tell an auditor? This post walks through what 29 CFR 1910.95 actually requires, why a lapsed date causes trouble even when the instrument was fine, and the order of operations Foster uses to put a program back on solid ground.

What does OSHA require for audiometer calibration?

Paragraph (h)(5) of the noise standard sets three checks on every audiometer used in a hearing conservation program. First, a functional check before each day's use, done by testing a person with known, stable hearing thresholds and listening for distorted or unwanted sounds. A deviation of 10 dB or more requires an acoustic calibration. Second, an acoustic calibration at least annually, following the mandatory procedure in Appendix E, which measures earphone output through a coupler and sound level meter at each test frequency from 500 to 6000 Hz and checks attenuator linearity. A deviation of 15 dB or more requires an exhaustive calibration. Third, an exhaustive calibration at least every two years against the ANSI S3.6 sections the standard lists. "At least annually" has no grace period written into it. If the last acoustic calibration was thirteen months ago, the instrument is past due on the plain words of the rule, whether or not anything about its output has changed.

Are hearing tests run on an overdue audiometer automatically invalid?

The standard does not say that. Nothing in 1910.95 declares an audiogram void because the calibration date lapsed. The problem is proof. Paragraph (m)(2)(ii)(D) requires every audiometric test record to include the date of the last acoustic or exhaustive calibration of the audiometer, so each audiogram taken during the lapse carries the evidence of the gap on its face. Paragraph (g)(7)(iii)(D) requires the employer to hand the reviewing audiologist or physician the audiometer calibration records when a problem audiogram is referred. That means the professional deciding whether a shift is real, and whether it is recordable, is looking at a test run on an instrument with no current calibration behind it. If the audiometer turns out to be within tolerance when it is finally checked, the tests are defensible but the record still shows the missed date. If it turns out to be out of tolerance, every test since the last good calibration is suspect.

What should you do first when you find the lapse?

Stop testing on that unit and get it calibrated before running another employee. When the provider checks it, ask that the readings be recorded as found, before any adjustment. Those as-found numbers are what tell you whether the tests already taken can stand. A unit that comes in within the Appendix E tolerances (3 dB from 500 through 3000 Hz, 4 dB at 4000 Hz, 5 dB at 6000 Hz) was very likely producing the levels it claimed during the lapse. A unit that comes in outside them was not, and the size and frequency of the error decide which results matter. Appendix E advises an exhaustive calibration when as-found readings fall outside those tolerances, and it requires one if the deviation is 15 dB or more. The professional supervisor still decides what to do with the tests already taken. Then write a short memo to the program file: when the lapse was found, which unit, the date range affected, the as-found result, and what was done about each test in that window. An auditor reads a documented correction very differently from an unexplained gap.

How do as-found results change what happens to past audiograms?

As-found resultWhat it suggestsTypical action on tests in the lapse window
Within Appendix E tolerance at all frequenciesOutput was accurate during the lapseKeep results, document the lapse and the as-found data in the program file
Outside tolerance but under 15 dBOutput error at one or more frequencies. Appendix E advises an exhaustive calibration at these deviations. It requires one at 15 dB or more.Have the professional supervisor review any STS, baseline, or borderline result at the affected frequencies
15 dB or more at any frequencyExhaustive calibration required, results unreliableRetest affected employees on a calibrated unit, prioritizing new baselines and any STS
Unit failed or could not be measuredNo way to know what was deliveredTreat the window as untested, schedule retests

The decisions about specific employees belong to the audiologist or physician reviewing the program, not to the calibration vendor. CAOHC's guidance to occupational hearing conservationists makes the same point: the tester is responsible for calibration and documentation, and quality control procedures should be worked out with the professional supervisor ahead of time, because follow-up on inaccurate results and retesting employees costs far more than the check.

Which audiograms from the lapse window matter most?

Three kinds of results carry the most weight. Baselines are first, because every future audiogram for that employee is compared against them, and a baseline taken on a unit reading several decibels low makes later results look like hearing loss. Standard threshold shifts are second. An STS triggers employee notification, refitting of hearing protectors, possible referral, and a recordability decision under 29 CFR 1904.10, so a shift that appeared only on a suspect instrument deserves a retest. The standard already allows a retest within 30 days to stand in for the annual audiogram when an STS shows up, which gives you a clean path. Results that came out normal on a unit that later checked within tolerance are the lowest priority. For anything involving the STS rules or baseline revision, Foster's baseline audiogram explainer covers what the reviewer is allowed to change and when.

Does an overdue calibration matter outside OSHA?

It can. Audiograms from a hearing conservation program often end up in workers' compensation claims years later. Under the Longshore and Harbor Workers' Compensation Act, for example, the Department of Labor's regulations already require that the audiometer used for testing be calibrated to current ANSI specifications, and in June 2026 the Office of Workers' Compensation Programs published a request for information on quality standards for hearing loss testing, including calibration requirements. A test record that shows a lapsed calibration date is easy for either side of a claim to challenge.

How Foster handles this

Foster calibrates audiometers on site with a reference sound level meter and coupler, so the unit is back in service the same day and the customer does not lose a testing week to shipping. Every certificate lists measured values at each frequency against tolerance, not a single pass line, which is what a reviewer needs when a lapse has to be explained. When we find a unit past due, we record the readings before touching any adjustment so the customer and their professional supervisor can make the call on prior tests with real numbers. Customers on our calibration service get a reminder ahead of each anniversary date, and a loaner when a unit has to go to the shop. For employers tracking calibration alongside baselines, retests, and other OSHA and DOT due dates, our Compliance Workspace keeps each requirement and its date in one place. If retests are needed, our mobile hearing testing team can run them on site in Greater Cincinnati and the tri-state. Request a quote for either.

For background on the regular schedule, see how often an audiometer needs calibration and signs an audiometer is drifting.

Frequently asked questions

Is there a grace period after the annual calibration date?

No. 1910.95(h)(5)(ii) says "at least annually." Many programs schedule calibration ten or eleven months out so a rescheduled visit still lands inside the year.

Can the daily functional check stand in for a missed annual calibration?

No. The daily check uses a person's thresholds and listening, and its action threshold is a 10 dB deviation. The annual Appendix E check measures actual sound pressure output with instruments and is a separate requirement.

Do we have to retest everyone tested during the lapse?

Not automatically. If the as-found calibration is within tolerance, results can usually stand with the lapse documented. Retests are driven by out-of-tolerance findings and by which results matter, as decided by the reviewing audiologist or physician.

Does an overdue calibration have to be reported to OSHA?

No report is required. The record of the lapse lives in your audiometric test records and calibration files, which must be available to employees and OSHA on request under 1910.95(m)(4).