Audiometer Calibration: At Least Once Per Year (OSHA, ANSI)

July 19, 2026 · 5 min read

By Jeff Schroeder, Foster Special Instruments

How often does an audiometer need calibration?

An audiometer used for OSHA hearing conservation testing needs calibration checks at three different intervals: a functional (biological) check before use each day it's used, an annual acoustic calibration at least once a year, and an exhaustive calibration at least every two years, or sooner when an annual check shows a deviation of 15 dB or more. Skipping any of these three isn't just a maintenance lapse; 29 CFR 1910.95 requires the audiometric test record to state the date of the instrument's last acoustic or exhaustive calibration, so a lapsed calibration invalidates the paperwork for every test run on that instrument. See Foster's audiometer calibration service for what is covered.

Need the annual calibration done? Foster’s audiometer calibration service is NIST-traceable to ANSI S3.6, mail-in or on-site, and every instrument ships back with a calibration certificate that lists the measured values at each frequency. See all of our audiometer calibration services for on-site, in-house, and mail-in options.

Does ANSI require audiometers to be calibrated at least once per year?

Yes, in practice. The audiometer must meet ANSI S3.6, and the schedule that keeps it in compliance comes from OSHA. Under 29 CFR 1910.95(h)(5), the functional check is done before each day's use, the acoustic calibration is checked at least annually per Appendix E, and an exhaustive calibration is performed at least every two years against the S3.6 specification. A deviation of 10 dB or more at the daily check requires an acoustic calibration, and a deviation of 15 dB or more at the annual check requires an exhaustive calibration. For other field instruments, see how ANSI requires field instruments be calibrated at least once per year or on the schedule that applies to each type.

What is a daily functional (biological) check?

Before an audiometer is used each testing day, the operator runs a quick biological check: a known-hearing person listens through the earphones at each frequency the device tests, at a low hearing level, and confirms the tones are audible, clear, and free of distortion, and that the attenuator dial and talk-back circuit work correctly. This isn't a formal calibration — it doesn't require reference equipment — but it catches gross failures like a dead earphone or a stuck dial before a technician tests dozens of employees on a broken instrument and has to retest all of them.

What is annual acoustic calibration?

Once a year, the audiometer's output at each test frequency is checked against a calibrated reference (a sound level meter and artificial ear or coupler) to confirm actual output matches the dial reading within the tolerance set by ANSI S3.6, the American National Standard for audiometers. This check catches drift that a functional check can't — an instrument can sound fine to the ear and still be off by several decibels at a specific frequency. Results and adjustments get logged, and that log date is what goes on the employee's audiometric record per 1910.95.

When is exhaustive calibration required?

Exhaustive calibration is a full electroacoustic evaluation against ANSI S3.6 specifications — checking not just output level but linearity, frequency accuracy, harmonic distortion, and attenuator function across the full range. Under 1910.95(h)(5)(iii) it must be performed at least every two years, and it is also required when the annual acoustic check finds a deviation of 15 dB or more. An instrument that fails exhaustive calibration standards gets taken out of service, not patched.

Check typeFrequencyWhat it verifies
Functional (biological)Each day of useAudible, clear tones at each frequency; dial and controls work
Annual acousticAt least once per yearOutput level accuracy vs. ANSI S3.6 tolerance at each frequency
ExhaustiveAt least every two years, or after a 15 dB deviationFull electroacoustic spec: level, linearity, distortion, attenuator

What does Foster's annual acoustic calibration visit involve?

A Foster cal tech brings a reference sound level meter and coupler to the customer's audiometer rather than shipping the unit out, so the instrument is back in service the same day. The tech runs each test frequency through the coupler, compares actual output to the dial reading at each step, and checks attenuator linearity across the range rather than just spot-checking one or two frequencies. Any frequency reading outside ANSI S3.6 tolerance gets adjusted on the spot if it's a routine drift; if the unit needs parts or shows distortion the field kit can't fix, it goes back to the shop and the customer gets a loaner audiometer so scheduled annual testing doesn't slip. The tech leaves a dated calibration certificate that lists every frequency checked, which is the document that has to match the date printed on each employee's audiometric record.

Using more than one type of instrument? See how often ANSI requires field instruments to be calibrated across audiometers, spirometers, and sound level meters.

What causes an audiometer to fail calibration that a daily check misses?

The most common failure a functional check can't catch is gradual output drift at one or two specific frequencies — the tones still sound clear to a listener, but the actual decibel output at, say, 4000 Hz has crept several dB off the dial reading over months of use. Earphone cushions that have hardened or lost their seal are another quiet failure mode: they don't sound obviously wrong but change the acoustic coupling enough to shift readings. Transport damage is the third common cause, especially for units that travel between job sites in a vehicle without a hard case — a dropped audiometer can pass a same-day functional check and still be out of spec on the next acoustic calibration.

Frequently Asked Questions

How often must audiometers be calibrated?

Check function before each day's use, check acoustic calibration at least once per year per Appendix E, and perform an exhaustive calibration at least every two years, all under 29 CFR 1910.95(h)(5).

What standard governs audiometer calibration tolerance?

ANSI S3.6, the American National Standard Specification for Audiometers, sets the accuracy tolerances that annual and exhaustive calibration are checked against.

Does a missed annual calibration invalidate audiograms taken that year?

It creates a recordkeeping and defensibility gap — 1910.95 requires the calibration date on the audiometric record, and testing on an instrument past its calibration due date is difficult to defend if a threshold shift determination is later challenged.

Can the same technician who tests employees also run the functional check?

Yes — the daily functional check is designed to be run by the testing technician as a quick pre-use verification, not a separate certified calibration event.

What should an audiometer calibration certificate include?

The certificate should name the instrument model and serial number, the calibration date, the reference equipment and its traceability, and the measured output at each frequency with the ANSI S3.6 tolerance, not just a pass stamp. Our guide to what a defensible calibration certificate must contain walks through each line, and Foster’s audiometer calibration certificate carries all of it.

Foster Special Instruments performs annual acoustic and exhaustive audiometer calibration on-site, in-house, or mail-in, and supports on-site hearing testing programs. For the full 1910.95 program requirements, see the OSHA hearing conservation checklist.